Accuracy comparison of a 16 and 64 multidetector-row computed tomography scanner to image small high-density structures

Invest Radiol. 2006 Nov;41(11):781-92. doi: 10.1097/01.rli.0000239319.86621.f5.

Abstract

Objectives: The accuracy in imaging small high-density structures is compared for 16 and 64 multidetector-row computed tomography (MDCT) scanners.

Materials and methods: Phantom experiments and different quantification methods are used to establish size measurement accuracy, object signal, and image noise, for both MDCT systems.

Results: At similar scanning doses, image noise is larger ( approximately 55%) for the 64 MDCT compared with the 16 MDCT, leading to lower signal-to-noise ratios ( approximately 28% for objects <2 mm). Object spread in the xy-plane is similar for both systems; while it is reduced along the z-axis (by approximately 0.18 mm) for the 64 MDCT. Measurement accuracy of the 64 MDCT is not significantly higher (P < 0.05) compared with the 16 MDCT when using a relative-threshold corresponding to 50% of the object maximum attenuation value. However, when using a fixed-threshold, interscanner and interprotocol measurement differences are statistically significant (eg, volume relative errors are reduced by approximately 17% on average for the 64 MDCT).

Conclusions: Measurement accuracy of the 16 and 64 MDCT scanners is not significantly different when using a 50% relative threshold. However, image noise is significantly larger for the 64 MDCT. Compared with a fixed-threshold based method, the 50% relative-threshold strongly reduces interscanner and interprotocol measurement dependency and improves accuracy.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Image Enhancement
  • Image Processing, Computer-Assisted
  • Polymethyl Methacrylate
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tomography Scanners, X-Ray Computed / standards*
  • Tomography, X-Ray Computed / instrumentation*

Substances

  • Polymethyl Methacrylate