Use of an inspiratory impedance threshold valve during chest compressions without assisted ventilation may result in hypoxaemia

Resuscitation. 2007 Mar;72(3):466-76. doi: 10.1016/j.resuscitation.2006.07.016. Epub 2006 Dec 5.

Abstract

Introduction: Although the concept of intermittent airway occlusion with the inspiratory impedance threshold valve (ITV) is a well-recognised strategy for improving efficiency of cardiopulmonary resuscitation (CPR), little is known about possible pulmonary side effects.

Methods: After a baseline chest CT-scan, 24 pigs with beating hearts undergoing apnoeic oxygenation received an injection of a contrast medium and were then assigned randomly to either active compression-decompression CPR with ITV (ACD ITV CPR), ACD CPR alone, or standard-CPR with ITV (standard-ITV CPR), or standard-CPR alone. After a maximum of 5 min of chest compressions or if oxygen saturation dropped below 70%, the experiment was stopped, haemodynamic variables and blood gas values were measured, and another CT-scan was performed; all animals underwent a 30 min recovery-period and a third subsequent CT-scan.

Results: At baseline arterial oxygen saturation by pulse oxymetry was 99% in all four groups; in both the ACD ITV CPR and the standard-ITV CPR groups, arterial oxygen saturation dropped below 70% within 126+/-9s, whereas chest compressions in all ACD CPR and standard-CPR pigs were performed over 5 min (P<0.001). Before stopping chest compressions arterial oxygen pressure decreased in the ACD ITV CPR group from 426+/-96 to 42+/-8 mmHg while it decreased in the ACD CPR group only from 415+/-116 to 197+/-127 mmHg (P<0.001 between groups); in the standard-ITV CPR group arterial oxygen partial pressure decreased from 427+/-109 to 34+/-5 mmHg while oxygen partial pressure decreased only from 467+/-44 to 144+/-98 mmHg in the standard-CPR group (P<0.004 between groups). After the second CT scan arterial oxygen partial pressure decreased further to 19+/-2 mmHg in the ACD ITV CPR versus 210+/-41 mmHg in the ACD CPR group; to 20+/-2 mmHg in the standard-ITV CPR versus 148+/-33 mmHg in the standard-CPR group. Lung-density values (Hounsfield units) were significantly higher in the ACD ITV CPR versus ACD CPR group (-134+/-54 versus -330+/-77) and standard-ITV CPR versus standard-CPR group (-98+/-50 versus -387+/-42). After a 30 min recovery-period, there were no significant differences in arterial oxygen partial pressure (ACD ITV CPR 275+/-110 mmHg versus ACD CPR 379+/-111 mmHg and standard-ITV CPR 265+/-138 mmHg versus standard CPR 367+/-55 mmHg). Furthermore, there were no differences in lung density values between groups after 30 min of recovery.

Conclusion: In this animal model with a beating heart, intermittent airway obstruction through an ITV combined with apnoeic oxygenation and without active ventilation resulted in hypoxaemia due to transiently impaired lung function.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Airway Resistance*
  • Animals
  • Cardiopulmonary Resuscitation / adverse effects*
  • Cardiopulmonary Resuscitation / instrumentation*
  • Disease Models, Animal
  • Equipment Design
  • Female
  • Heart Arrest / therapy*
  • Hypoxia / diagnostic imaging
  • Hypoxia / etiology*
  • Male
  • Swine
  • Tomography, X-Ray Computed
  • Treatment Outcome