Cytomegalovirus quantification in plasma by an automated real-time PCR assay

J Clin Virol. 2007 Apr;38(4):298-303. doi: 10.1016/j.jcv.2007.01.003. Epub 2007 Feb 23.

Abstract

Background: Sensitive quantitation of cytomegalovirus (CMV) DNA in blood is helpful for the diagnosis of CMV infection or reactivation and the monitoring of transplanted patients.

Objectives: We compared a new PCR assay coupled with an automated extraction system (CMV real-time PCR, Abbott Molecular, Des Plaines, IL, USA) to a previously validated method (ultrasensitive Cobas Amplicor CMV DNA Monitor, Roche Molecular, Indianapolis, IN, USA).

Results: Using limiting dilutions of CMV DNA positive plasma, the two assays had a similar detection threshold ranging between 20 and 45 copies/ml. Coefficients of variation of CMV real-time PCR assay varied from 1 to 12% for CMV DNA levels between 10,000 and 20 copies/ml. Viral loads assessed by the two methods on 179 clinical samples showed an overall concordance of 89% and an excellent correlation (R=0.94). Discrepancies were only observed for samples with low CMV DNA levels (<300 copies/ml); 18 samples were positive by CMV real-time PCR only, and 2 samples by ultrasensitive Cobas CMV only. Values obtained by CMV real-time PCR were on average 0.4 log higher than those of ultrasensitive Cobas CMV. Successive samples of transplanted patients with evidence of CMV infection/reactivation revealed that CMV real-time PCR assay was positive earlier and for a longer period of time after treatment initiation.

Conclusions: Although both assays had similar analytical performances, the CMV real-time PCR assay has the advantages of automated extraction and higher dynamic range, and shows a trend for an improved sensitivity that might impact on clinical decisions.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Adult
  • Analysis of Variance
  • Cytomegalovirus / genetics*
  • DNA, Viral / blood*
  • Female
  • Humans
  • Plasma / virology
  • Polymerase Chain Reaction / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Viral Load / methods*

Substances

  • DNA, Viral