4 T Actively detuneable double-tuned 1H/31P head volume coil and four-channel 31P phased array for human brain spectroscopy

J Magn Reson. 2007 Jun;186(2):341-6. doi: 10.1016/j.jmr.2007.03.001. Epub 2007 Mar 3.

Abstract

Typically 31P in vivo magnetic resonance spectroscopic studies are limited by SNR considerations. Although phased arrays can improve the SNR; to date 31P phased arrays for high-field systems have not been combined with 31P volume transmit coils. Additionally, to provide anatomical reference for the 31P studies, without removal of the coil or patient from the magnet, double-tuning (31P/1H) of the volume coil is required. In this work we describe a series of methods for active detuning and decoupling enabling use of phased arrays with double-tuned volume coils. To demonstrate these principles we have built and characterized an actively detuneable 31P/1H TEM volume transmit/four-channel 31P phased array for 4 T magnetic resonance spectroscopic imaging (MRSI) of the human brain. The coil can be used either in volume-transmit/array-receive mode or in TEM transmit/receive mode with the array detuned. Threefold SNR improvement was obtained at the periphery of the brain using the phased array as compared to the volume coil.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Artifacts
  • Brain / anatomy & histology*
  • Humans
  • Magnetic Resonance Imaging / instrumentation*
  • Magnetic Resonance Spectroscopy / instrumentation*
  • Phosphorus Isotopes
  • Protons*
  • Radio Waves