Mechanical performance of hybrid Ilizarov frames containing full-threaded Schanz screws

Orthopedics. 2007 Apr;30(4):304-7. doi: 10.3928/01477447-20070401-09.

Abstract

Modification of circular Ilizarov frames is necessary to overcome the difficulties in pin positioning due to unfavorable anatomic site or local soft-tissue conditions. Hybrid frame configurations consisting of half pins or full-threaded schanz screws are widely used in clinical practice. This study compared the mechanical performance of hybrid frames and a standard system. One standard and five modified hybrid systems were tested under axial compression, four-point bending, and torsional forces. Systems modified with full-threaded schanz screws showed a higher stiffness than half pin modifications and exhibited a similar mechanical performance of a standard system.

MeSH terms

  • Bone Screws*
  • Equipment Design
  • Humans
  • Ilizarov Technique / instrumentation*
  • Materials Testing*
  • Torsion Abnormality