A 2 m inelastic X-ray scattering spectrometer at CMC-XOR, Advanced Photon Source

J Synchrotron Radiat. 2007 Jul;14(Pt 4):361-5. doi: 10.1107/S0909049507018006. Epub 2007 Jun 14.

Abstract

The design and commissioning of an inelastic X-ray scattering instrument at CMC-XOR at the Advanced Photon Source is reported. The instrument features a 2 m vertical-scattering arm with a novel counterweight design to reduce the twisting moment as the arm is moved in the scattering plane. A Ge(733) spherical analyzer was fabricated and an overall resolution of 118 meV (FWHM) was obtained with a Si(444) monochromator and a Si(111) pre-monochromator. Early results from a representative cuprate, La(2)CuO(4), are reported.