We performed two image reading experiments in order to investigate the diagnostic capability of I-123 IMP SPECT obtained by the ring type SPECT scanner in cerebro-vascular disease. Fourteen physicians diagnosed SPECT images of 55 cases with reference to clinical neurological informations, first without brain XCT images and second with XCT images. Each physician detected perfusion defects and redistributions of I-123 IMP and assigned a confidence level of abnormality for these SPECT findings by means of five rating method. From results obtained by ROC analysis, we concluded as follows: (1) Generally, I-123 IMP SPECT is a stable diagnostic modality in the diagnosis of cerebro-vascular disease and the image reading of XCT had no effects on the diagnosis of SPECT on the whole of physician, (2) However, there were unnegligible differences among individuals in the detectability of findings and the effect of XCT image reading, (3) Detectability of redistribution of I-123 IMP was lower than that of perfusion defect and inter-observer variation in the diagnostic performance for redistribution was larger than that of perfusion defect. The results suggest that it is necessary to standardize diagnostic criteria among physicians for redistribution of I-123 IMP.