Accuracy of implant placement based on pre-surgical planning of three-dimensional cone-beam images: a pilot study

J Clin Periodontol. 2007 Sep;34(9):816-21. doi: 10.1111/j.1600-051X.2007.01110.x.

Abstract

Aim: To evaluate the precision of transfer of a computer-based three-dimensional (3D) planning, using re-formatted cone-beam images, for oral implant placement in partially edentulous jaws.

Material and methods: Four formalin-fixed cadaver jaws were imaged in a 3D Accuitomo FPD cone-beam computed tomography (CT). Data were used to produce an accurate implant planning with a transfer to surgery by means of stereolithographic drill guides. Pre-operative cone-beam CT images were subsequently matched with post-operative ones to calculate the deviation between planned and installed implants.

Results: Placed implants (length: 10-15 mm) showed an average angular deviation of 2 degrees (SD: 0.8, range: 0.7-4.0 degrees ) as compared with the planning, while the mean linear deviation was 1.1 mm (SD: 0.7 mm, range 0.3-2.3 mm) at the hex and 2.0 mm (SD: 0.7 mm, range 0.7-2.4 mm) at the tip.

Conclusions: Cone-beam images could be used for implant planning, taking into account a maximal 4 degrees angular and 2.4 mm linear deviation at the apical tip.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cadaver
  • Dental Implantation, Endosseous / instrumentation
  • Dental Implantation, Endosseous / methods*
  • Dental Implants*
  • Humans
  • Image Processing, Computer-Assisted / methods
  • Imaging, Three-Dimensional / methods*
  • Jaw, Edentulous, Partially / diagnostic imaging
  • Jaw, Edentulous, Partially / surgery*
  • Methacrylates
  • Patient Care Planning*
  • Pilot Projects
  • Surgery, Computer-Assisted
  • Tomography, X-Ray Computed / methods*

Substances

  • Dental Implants
  • Methacrylates