Analysis of adenosine by RP-HPLC method and its application to the study of adenosine kinase kinetics

J Sep Sci. 2007 Oct;30(15):2473-9. doi: 10.1002/jssc.200700194.

Abstract

An RP-HPLC method for the analysis of adenosine (ADO) has been developed and validated. In the present study, we report an RP-HPLC-based method with modifications of mobile phase and shorter retention time that substantially improved the efficiency of ADO analysis. The HPLC separation of the ADO was achieved on a C18 column, using a mobile phase consisting of water, containing 7% v/v ACN, at a flow rate of 0.8 mL/min. The column effluent was monitored by UV detection at 260 nm. A linear response was achieved over the concentration range of 0.25-100.00 micromol/L. The analytical method inter- and intra-run accuracy and precision were better than +/- 15%. The LOQ was 0.25 micromol/L, with ADO detection in the range of 6.25 pmol per sample. The method has been applied to the study of adenosine kinase (AK) kinetics.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adenosine / analysis*
  • Adenosine Kinase / metabolism*
  • Animals
  • Chemistry Techniques, Analytical / methods*
  • Chromatography / methods
  • Chromatography, High Pressure Liquid / methods*
  • Kinetics
  • Mice
  • Quality Control
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrophotometry, Ultraviolet
  • Time Factors
  • Water / chemistry*

Substances

  • Water
  • Adenosine Kinase
  • Adenosine