Optimization and validation of a high-performance liquid chromatographic method with UV detection for the determination of pyrrole-imidazole polyamides in rat plasma

J Chromatogr B Analyt Technol Biomed Life Sci. 2007 Nov 15;859(2):272-5. doi: 10.1016/j.jchromb.2007.09.032. Epub 2007 Oct 5.

Abstract

A simple and sensitive high-performance liquid chromatography (HPLC) method utilizing UV detection was developed for the determination of plasma pyrrole (Py)-imidazole (Im) polyamides in rats and applied to the pharmacokinetic study of compounds. After deproteinization of plasma with methanol, Py-Im polyamides were analyzed with a reversed-phase TSK-GEL ODS-80TM (4.6 mmx15.0 cm TOSOH Co., Japan) column maintained at 40 degrees C. The mobile phase solvent A was 0.1% acetic acid and the solvent B was HPLC-grade acetonitrile (0-10 min, A: 100-20%, B: 0-80% linear gradient; 10-15 min, A: 40%, B: 60%). The flow rate was 1.0 ml/min. The detection wavelength was set at 310 nm. The method was used to determine the plasma concentration time profiles of Py-Im polyamides after intravenous injection.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Animals
  • Chromatography, High Pressure Liquid / methods*
  • Imidazoles / blood*
  • Male
  • Nylons / analysis*
  • Pyrroles / blood*
  • Rats
  • Rats, Wistar
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Ultraviolet Rays

Substances

  • Imidazoles
  • Nylons
  • Pyrroles