A 3D inner-volume fast spin echo (3D IV-FSE) sequence was developed for ECG-gated, black-blood, T1- and T2-weighted vessel wall imaging of peripheral vein bypass grafts (PVBG). The sequence utilizes nonselective refocusing excitations to minimize echo spacings and a highly selective IV excitation scheme to minimize the need for oversampling of z-encode slice selections. The method was tested in eight PVBG patients who also underwent 2D FSE graft imaging. High-quality 3D imaging was achieved in all subjects, with significant spatial resolution and volume coverage gains compared to the more conventional 2D FSE sequences normalized for signal-to-noise ratios (SNRs) and scan times. Compared to previously proposed 3D IV-FSE methods, nonselective refocusing resulted in a more than 20% FSE echo train sampling efficiency increase while the use of highly selective IV excitation resulted in a 30% improvement in slice oversampling efficiency.
(c) 2008 Wiley-Liss, Inc.