High-resolution peripheral vein bypass graft wall studies using high sampling efficiency inner volume 3D FSE

Magn Reson Med. 2008 Mar;59(3):650-4. doi: 10.1002/mrm.21359.

Abstract

A 3D inner-volume fast spin echo (3D IV-FSE) sequence was developed for ECG-gated, black-blood, T1- and T2-weighted vessel wall imaging of peripheral vein bypass grafts (PVBG). The sequence utilizes nonselective refocusing excitations to minimize echo spacings and a highly selective IV excitation scheme to minimize the need for oversampling of z-encode slice selections. The method was tested in eight PVBG patients who also underwent 2D FSE graft imaging. High-quality 3D imaging was achieved in all subjects, with significant spatial resolution and volume coverage gains compared to the more conventional 2D FSE sequences normalized for signal-to-noise ratios (SNRs) and scan times. Compared to previously proposed 3D IV-FSE methods, nonselective refocusing resulted in a more than 20% FSE echo train sampling efficiency increase while the use of highly selective IV excitation resulted in a 30% improvement in slice oversampling efficiency.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Electrocardiography
  • Female
  • Humans
  • Imaging, Three-Dimensional*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Peripheral Vascular Diseases / surgery*
  • Vascular Surgical Procedures / methods*
  • Veins / transplantation*