Visualization of macromolecular complexes using cryo-electron microscopy with FEI Tecnai transmission electron microscopes

Nat Protoc. 2008;3(2):330-9. doi: 10.1038/nprot.2007.474.

Abstract

This protocol details the steps used for visualizing the frozen-hydrated grids as prepared following the accompanying protocol entitled 'Preparation of macromolecular complexes for visualization using cryo-electron microscopy.' This protocol describes how to transfer the grid to the microscope using a standard cryo-transfer holder or, alternatively, using a cryo-cartridge loading system, and how to collect low-dose data using an FEI Tecnai transmission electron microscope. This protocol also summarizes and compares the various options that are available in data collection for three-dimensional (3D) single-particle reconstruction. These options include microscope settings, choice of detectors and data collection strategies both in situations where a 3D reference is available and in the absence of such a reference (random-conical and common lines).

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Cryoelectron Microscopy / methods*
  • Image Processing, Computer-Assisted
  • Imaging, Three-Dimensional / methods*
  • Macromolecular Substances / analysis*
  • Microscopy, Electron, Transmission / methods*

Substances

  • Macromolecular Substances