Magn Reson Med. 2008 Apr;59(4):686-91. doi: 10.1002/mrm.21401.

Abstract

Rapid MR imaging using the stimulated echo acquisition mode (STEAM) technique yields single-shot images without any sensitivity to resonance offset effects. However, the absence of susceptibility-induced signal voids or geometric distortions is at the expense of a somewhat lower signal-to-noise ratio than EPI. As a consequence, the achievable spatial resolution is limited when using conventional Fourier encoding. To overcome the problem, this study combined single-shot STEAM MRI with radial encoding. This approach exploits the efficient undersampling properties of radial trajectories with use of a previously developed iterative image reconstruction method that compensates for the incomplete data by incorporating a priori knowledge. Experimental results for a phantom and human brain in vivo demonstrate that radial single-shot STEAM MRI may exceed the resolution obtainable by a comparable Cartesian acquisition by a factor of four.

MeSH terms

  • Algorithms*
  • Brain / anatomy & histology*
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / instrumentation
  • Magnetic Resonance Imaging / methods*
  • Numerical Analysis, Computer-Assisted
  • Phantoms, Imaging
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Signal Processing, Computer-Assisted*