Statpac 2 glaucoma change probability

Arch Ophthalmol. 1991 Dec;109(12):1690-2. doi: 10.1001/archopht.1991.01080120074029.

Abstract

We compared the traditional intuitive criterion (fixed at 5 dB) with the variable Statpac 2 criterion of the Humphrey Field Analyzer for identifying visual deterioration at individual points within a visual field. The separation of depressed from stable points in 123 follow-up fields of 17 patients, when all points within a field were considered, was reasonably equivalent between the two methods (kappa = 0.55 +/- 0.027 [+/- SE]). Centrally located points within the field demonstrated a stronger agreement (kappa = 0.68 +/- 0.028), because the fixed criterion more often labeled the edge points as having progressed than did the Statpac 2 algorithm. The glaucoma change probability printout is more simply obtained than the manual comparison required for intuitive interpretation with a fixed criterion, and it is potentially more accurate because of its reliance on a database that permits a variable criterion to be applied.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Aged
  • Diagnosis, Computer-Assisted*
  • Female
  • Follow-Up Studies
  • Glaucoma / physiopathology*
  • Humans
  • Male
  • Middle Aged
  • Probability
  • Visual Field Tests / methods*