[A simulative biomechanical experiment on different position of none-cement acetabular components influencing the load distribution around acetabulum]

Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2008 Feb;25(1):97-100.
[Article in Chinese]

Abstract

In the present study, a three dimensional finite-element model of the human pelvic was reconstructed, and then, under different acetabular component position (the abduction angle ranges from 30 degrees to 70 degrees and the anteversion ranges from 5 degrees to 30degrees) the load distribution around the acetabular was evaluated by the computer biomechanical analysis program (Solidworks). Through the obtained load distribution results, the most even and reasonable range of the distribution was selected; therefore the safe range of the acetabular component implantation can be validated from the biomechanics aspect.