Liquid chromatography tandem mass spectrometry method for the quantification of clonidine with LLOQ of 10 pg/mL in human plasma

Biomed Chromatogr. 2008 Sep;22(9):992-1000. doi: 10.1002/bmc.1018.

Abstract

A sensitive high-performance liquid chromatography-positive ion electrospray tandem mass spectrometry method was developed and validated for the quantification of clonidine in human plasma. Following liquid-liquid extraction, the analytes were separated using an isocratic mobile phase on a reverse-phase column and analyzed by MS/MS in the multiple reaction monitoring mode using the respective [M + H](+) ions, m/z 230 to 44 for clonidine and m/z 254 to 44 for the internal standard. The assay exhibited a linear dynamic range of 10-2000 pg/mL for clonidine in human plasma. The lower limit of quantification was 10 pg/mL with a relative standard deviation of less than 6.8%. Acceptable precision and accuracy were obtained for concentrations over the standard curve range. A run time of 2.5 min for each sample made it possible to analyze more than 250 human plasma samples per day. The validated method was successfully used to analyze human plasma samples for application in pharmacokinetic studies.

Publication types

  • Validation Study

MeSH terms

  • Antihypertensive Agents / blood*
  • Antihypertensive Agents / chemistry
  • Chromatography, High Pressure Liquid / methods*
  • Clonidine / blood*
  • Clonidine / chemistry
  • Humans
  • Sensitivity and Specificity
  • Tandem Mass Spectrometry / methods*