Measurement of edge density profiles of Large Helical Device plasmas using an ultrashort-pulse reflectometer

Rev Sci Instrum. 2008 May;79(5):056106. doi: 10.1063/1.2917579.

Abstract

We report here on the application of an ultrashort-pulse reflectometer (USPR) to Large Helical Device in National Institute for Fusion Science. An impulse with picosecond pulse width is used as a source in an USPR. Since the bandwidth of a source is inversely related to the pulse width, we can utilize the frequency range of microwave to millimeter-wave by using wide band transmission lines. The density profiles can be reconstructed by collecting time-of-flight signal of each frequency component of an impulse reflected from each cutoff layer. Remote control system using super science information network has been introduced to the present USPR system.