Semi-automatic measurement of the airway dimension by computed tomography using the full-width-half-maximum method: a study on the measurement accuracy according to the CT parameters and size of the airway

Korean J Radiol. 2008 May-Jun;9(3):226-35. doi: 10.3348/kjr.2008.9.3.226.

Abstract

Objective: To assess the influence of variable factors such as the size of the airway and the CT imaging parameters such as the reconstruction kernel, field-of-view (FOV), and slice thickness on the automatic measurement of airway dimension.

Materials and methods: An airway phantom was fabricated that contained eleven poly-acryl tubes of various lumen diameters and wall thicknesses. The measured density of the poly-acryl wall was 150 HU, and the measured density of the airspace filled with polyurethane foam was -900 HU. CT images were obtained using a 16-MDCT (multidetector CT) scanner and were reconstructed with various reconstruction kernels, thicknesses and FOV. The luminal radius and wall thickness were measured using in-house software based on the full-width-half-maximum method. The measured values as determined by CT and the actual dimensions of the tubes were compared.

Results: Measurements were most accurate on images reconstructed with use of a standard kernel (mean error: -0.03 +/- 0.21 mm for wall thickness and -0.12 +/- 0.11 mm for the luminal radius). There was no significant difference in accuracy among images with the use of variable slice thicknesses or a variable FOV. Below a 1-mm threshold, the measurement failed to represent the change of the real dimensions.

Conclusion: Measurement accuracy was strongly influenced by the specific reconstruction kernel utilized. For accurate measurement, standardization of the imaging protocol and selection of the appropriate anatomic level are essential.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cone-Beam Computed Tomography / methods*
  • Feasibility Studies
  • Phantoms, Imaging
  • Respiratory System / anatomy & histology*