Background: The Medtronic Sprint Fidelis implantable cardioverter-defibrillator high-voltage lead is prone to fracture. The October 2007 safety advisory recommended lead impedance monitoring to aid in identifying lead fractures.
Objective: The aim of this single-center study was to examine the effectiveness of impedance monitoring for detecting Sprint Fidelis lead failures before they caused adverse clinical events such as inappropriate shocks.
Methods: Impedance and sensing information were acquired during routine clinic and CareLink follow-up and at the time of lead failure using the Patient Alert, sensing integrity counter, nonsustained episode, and electrogram features in Medtronic pulse generators.
Results: Between September 2004 and February 2008, 17 of 514 Sprint Fidelis leads (3.3%) followed up at our center failed between 11 and 35 months after implantation (mean 23.0 +/- 8.0 months). Fifteen of these failures (88%) were caused by pace-sense conductor fractures, and 2 (12%) were caused by high-voltage conductor defects. Twelve of 15 patients (80%) with pace-sense conductor fractures received inappropriate shocks; of these, 4 had no significant increase in lead impedance before they were shocked, 2 were shocked <3 hours after their lead impedances exceeded the 1,000 Omega audible alert threshold, and 2 patients did not hear the alarm. All pace-sense conductor failures whose sensing function could be evaluated (13 of 15) had oversensing based on stored data, and oversensing usually (11 of 13) preceded impedances changes.
Conclusion: Impedance monitoring did not prevent inappropriate shocks in two-thirds of our patients. Thus, pace-sense conductor impedance monitoring as currently implemented does not reliably forewarn patients of a lead malfunction. Consequently, patients who have Sprint Fidelis leads remain at risk for adverse clinical events associated with pace-sense conductor fracture.