Application of a microcoil probe head in NMR analysis of chemicals related to the chemical weapons convention

Anal Chem. 2008 Jul 15;80(14):5556-64. doi: 10.1021/ac800623p. Epub 2008 Jun 12.

Abstract

A 1.7-mm microcoil probe head was tested in the analysis of organophosphorus compounds related to the Chemical Weapons Convention. The microcoil probe head demonstrated a high mass sensitivity in the detection of traces of organophosphorus compounds in samples. Methylphosphonic acid, the common secondary degradation product of sarin, soman, and VX, was detected at level 50 ng (0.52 nmol) from a 30-microL water sample using proton-observed experiments. Direct phosphorus observation of methylphosphonic acid with (31)P{(1)H} NMR experiment was feasible at the 400-ng (4.17 nmol) level. Application of the microcoil probe head in the spiked sample analysis was studied with a test water sample containing 2-10 microg/mL of three organophosphorus compounds. High-quality (1)H NMR, (31)P{(1)H} NMR, 2D (1)H-(31)P fast-HMQC, and 2D TOCSY spectra were obtained in 3 h from the concentrated 1.7-mm NMR sample prepared from 1 mL of the water solution. Furthermore, a 2D (1)H-(13)C fast-HMQC spectrum with sufficient quality was possible to measure in 5 h. The microcoil probe head demonstrated a considerable sensitivity improvement and reduction of measurement times for the NMR spectroscopy in identification of chemicals related to the Chemical Weapons Convention.