Introduction: Computed tomography (CT) is a widespread and highly precise technique working in the energy range around 50-100 keV. For radiotherapy, however, the MeV energy range enables a better dose distribution. This gap between diagnosis and therapy can be overcome by the use of a modified CT machine in combination with heavy elements targeted to the tumour and used as photoelectric radiation enhancer.
Materials and methods: The experimental setup consists of an X-ray optical module mounted at the exit of the X-ray tube of a clinical CT. The module converts the standard fan-shaped beam into a high intensity, monochromatized and focused beam. The radiation was characterized using an energy-dispersive detection system calibrated by synchrotron radiation and gel dosimetry. The photoelectric radiation enhancement for different elements was calculated and experimentally verified.
Results: The X-ray optical module filters selectively the energy of the tungsten K alpha-emission line (59.3 keV) with a full width at half maximum (FWHM) of 5 keV and focused the radiation onto a focal spot which coincides with the isocentre of the gantry. This results in a steep dose gradient at the centre of rotation qualified for locoregional radiation therapy. The photon energy of the quasi-monochromatic radiation agrees with the energy range of maximal photoelectric dose enhancement for gadolinium and iodine.
Conclusion: An additional X-ray optical module optimized for targeted therapy and photoelectric dose enhancement allows the combination of diagnosis and radiotherapy on a clinical CT.