[Detecting sleep apnea/hypopnea events with a wearable respiratory inductive plethysmograph system]

Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2008 Apr;25(2):318-22.
[Article in Chinese]

Abstract

The wearable respiratory inductive plethysmograph(RIP) system is a non-intrusive respiratory monitoring system. Sleep monitoring was performed on 9 human subjects suspected of having sleep apnea hypersomnolence syndrome (SAHS) and 7 healthy volunteers using both the wearable RIP system and the conventional polysomnography(PSG), and the sensitivity and specificity of the wearable RIP system were analyzed by comparison with the PSG results. According to the characteristic of the wearable RIP system in detecting sleep apnea/hypopnea event, the diagnostic criteria of sleep apnea/hypopnea event were put forward. All subjects with SAHS diagnosed by the wearable RIP were confirmed by PSG,the sensitivity and specificity of the wearable RIP system for detecting sleep apnea/hypopnea events were 99.0% and 94.6% respectively. The wearable RIP system can be used reliably in detecting sleep apnea/hypopnea events. This system can be used at home for detecting the sleep apnea/hypopnea events non-intrusively.

Publication types

  • English Abstract

MeSH terms

  • Equipment Design
  • Humans
  • Plethysmography / instrumentation*
  • Plethysmography / methods
  • Polysomnography / instrumentation*
  • Polysomnography / methods
  • Sleep Apnea Syndromes / diagnosis*
  • Sleep Apnea Syndromes / physiopathology*