A methodology for validating a 3D imaging modality for brain AVM delineation: application to 3DRA

Comput Med Imaging Graph. 2008 Oct;32(7):544-53. doi: 10.1016/j.compmedimag.2008.06.003. Epub 2008 Jul 18.

Abstract

A general methodology is described to validate a 3D imaging modality with respect to 2D digital subtracted angiography (DSA) for brain AVMs (BAVM) delineation. It relies on the assessment of the statistical compatibility of the radiosurgical target delineated in 3D with its delineations in 2D. This methodology is demonstrated through a preliminary evaluation of 3D rotational angiography (3DRA). Generally speaking, BAVM delineation cannot be performed on 3DRA alone. However, in our study, 3DRA showed similar performances to DSA for rather easy cases, and even better for three patients. Conversely, three problematic cases are identified and discussed.

Publication types

  • Validation Study

MeSH terms

  • Algorithms*
  • Angiography, Digital Subtraction / methods*
  • Artificial Intelligence*
  • Humans
  • Imaging, Three-Dimensional / methods*
  • Intracranial Arteriovenous Malformations / diagnostic imaging*
  • Pattern Recognition, Automated / methods*
  • Radiographic Image Enhancement / methods
  • Radiographic Image Interpretation, Computer-Assisted / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity