Assessment of knee alignment after bicompartmental knee arthroplasty

J Arthroplasty. 2009 Oct;24(7):1111-4. doi: 10.1016/j.arth.2008.07.006. Epub 2008 Oct 11.

Abstract

A key factor in successful knee arthroplasty is restoration of correct knee alignment. Although traditional total knee arthroplasty effectively restores alignment, it is unclear whether similar results can be consistently obtained with more conservative procedures. The purpose of this study was to assess alignment after implantation of the Journey Deuce Bi-Compartmental Knee System (Smith & Nephew, Inc, Memphis, Tenn). Mechanical axis was assessed in 137 cases. Preoperatively, 77 (56%) of 137 patients had a mechanical axis passing far medial to the center of the knee, consistent with varus deformity. Postoperatively, 130 (95%) of 137 patients had a mechanical axis passing through the center of the knee. These results suggest that the Journey Deuce Bi-Compartmental Knee System supports correction of varus deformity, previously thought to require total knee arthroplasty or osteotomy.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Arthroplasty, Replacement, Knee / instrumentation*
  • Arthroplasty, Replacement, Knee / methods*
  • Biomechanical Phenomena
  • Bone Malalignment / diagnostic imaging*
  • Bone Malalignment / prevention & control
  • Equipment and Supplies
  • Female
  • Follow-Up Studies
  • Humans
  • Knee Joint / diagnostic imaging*
  • Knee Joint / surgery*
  • Knee Prosthesis*
  • Male
  • Middle Aged
  • Osteoarthritis, Knee / surgery
  • Outcome Assessment, Health Care
  • Radiography
  • Retrospective Studies