Ablation monitoring with elastography: 2D in-vivo and 3D ex-vivo studies

Med Image Comput Comput Assist Interv. 2008;11(Pt 2):458-66. doi: 10.1007/978-3-540-85990-1_55.

Abstract

The clinical feasibility of 2D elastography methods is hindered by the requirement that the operator avoid out-of-plane motion of the ultrasound image during palpation, and also by the lack of volumetric elastography measurements. In this paper, we develop and evaluate a 3D elastography method operating on volumetric data acquired from a 3D probe. Our method is based on minimizing a cost function using dynamic programming (DP). The cost function incorporates similarity of echo amplitudes and displacement continuity. We present, to the best of our knowledge, the first in-vivo patient studies of monitoring liver ablation with freehand DP elastography. The thermal lesion was not discernable in the B-mode image but it was clearly visible in the strain image as well as in validation CT. We also present 3D strain images from thermal lesions in ex-vivo ablation. Good agreement was observed between strain images, CT and gross pathology.