Safety and efficacy of transvenous high-voltage implantable cardioverter-defibrillator leads in high-risk hypertrophic cardiomyopathy patients

Heart Rhythm. 2008 Nov;5(11):1517-22. doi: 10.1016/j.hrthm.2008.08.021. Epub 2008 Aug 28.

Abstract

Background: The implantable cardioverter-defibrillator (ICD) prevents sudden cardiac death in high-risk patients with hypertrophic cardiomyopathy (HCM). However, recently concerns have been raised regarding the performance of transvenous high-voltage ICD leads (HVL) in this population.

Objective: The purpose of this retrospective, multicenter study was to assess the safety and efficacy of HVL in high-risk HCM patients.

Methods: The study population consisted of HCM patients who received HVL ICDs and were subsequently followed at seven centers in the United States. Kaplan-Meier survival rates were calculated for HVL and patients. HVL failure was a malfunction caused by a noniatrogenic defect.

Results: Between 1992 and 2007, 324 HCM patients (mean age 47 +/- 16 years) received 343 HVL from three major manufacturers. The average HVL implant duration was 3.3 +/- 2.8 years. Overall, the HVL failure rate was 1.4%/year. However, two models (Sprint Fidelis and Transvene, Medtronic, Inc.) accounted for 60% of HVL failures. Survival probabilities for HVL and patients at 10 years were 93% and 91%, respectively. No deaths or serious injuries were reported, although inappropriate shocks occurred in 12% of cases.

Conclusions: This multicenter experience shows that HVLs are safe and effective in high-risk HCM patients. However, differences in failure rates were found between lead models.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Cardiomyopathy, Hypertrophic / therapy*
  • Defibrillators, Implantable*
  • Equipment Failure
  • Equipment Safety
  • Female
  • Humans
  • Male
  • Middle Aged
  • Retrospective Studies
  • Risk Factors
  • Survival Rate
  • Treatment Outcome