Objectives: To describe our experience of treating juxtarenal (JRAAA's <4mm neck) and thoracoabdominal aortic aneurysms (TAAA's) using fenestrated and branched stent graft technology.
Design: Prospective single centre experience.
Methods: Since 2005, 29 fenestrated/branched procedures have been performed. 15 patients are studied with JRAAAs (n=7; median neck length 0mm (IQR 0-3.8)) or TAAAs (type I (n=2), III (n=2), IV (n=4)). ASA grade III in 12/15. Maximum diameter of aneurysm 64 mm (56-74 mm). Aneurysms were excluded using covered stents or branches from the main body to patent visceral vessels (40 target vessels total). Pre-operative and follow-up CT scans (1, 3, and 12 months) were analysed by a single Vascular Interventional Radiologist.
Results: Technical success for cannulation and stenting of target vessels was 98%. In-hospital mortality was 0%. One patient underwent conversion to open repair. Five had major complications including one paraplegia (type III TAAA) with subsequent recovery. Median length of stay was 9 days (IQR 7-18.75). At a median follow-up of 12 months (9-14), CT confirmed 36/37 (97%) target vessels remain patent. Sac size increased >5 mm in one patient only. There were no type I endoleaks, three type II endoleaks (one embolised, two under surveillance) and three type III endoleaks (two successfully treated percutaneously, one aneurysm ruptured 18 months after endografting and died).
Conclusion: In selected patients, fenestrated and branched stents appear to be a safe and effective alternative to surgery for juxtarenal and thoracoabdominal aneurysms. The complication and mortality rates are low. The long-term durability of this procedure, however, needs to be proven.