Validation of an echocardiographic multiparametric strategy to increase responders patients after cardiac resynchronization: a multicentre study

Eur Heart J. 2009 Dec;30(23):2880-7. doi: 10.1093/eurheartj/ehn582. Epub 2009 Jan 9.

Abstract

Aims: We sought to develop and validate a multiparametric algorithm by applying previously validated criteria to predict cardiac resynchronization therapy (CRT) response in a multicentre study. Thirty per cent of patients treated by CRT fail to respond to the treatment. Although dyssynchrony by echocardiography has been used to improve the selection of patients, the complexity of myocardial contraction has generated a moderate improvement using any of several individual parameters.

Methods and results: Two hundred end-stage heart failure patients [NYHA 3-4 and left ventricular ejection fraction (LVEF)<35%] with QRS>120 ms were included. Echocardiography analysis focused on the following parameters: atrioventricular dyssynchrony, interventricular dyssynchrony, and intraventricular dyssynchrony that integrated radial (PSAX M-mode) and longitudinal [tissue Doppler imaging (TDI)] evaluations for spatial (wall to wall) and temporal (wall end-systole to mitral valve opening) dyssynchrony diagnosis. Following CRT implantation, patients were monitored for 6 months with functional and echo evaluations defining responders by a 15% reduction in end-systolic volume. Mean QRS duration and LVEF were 152 +/- 17 ms and 25 +/- 8%. There was a CRT response in 57% of patients, independent of QRS width. Mean prevalence of positive criteria was 34 +/- 8%. Feasibility and variability averages were 81 +/- 20% and 9 +/- 4%. In a single parametric approach, ranges of sensitivities and specificities were 18-65% and 45-84% with a mean of 41% and 66%. A multiparametric approach by focusing on criteria combination decreased the mean rate of false-positive results to 14 +/- 12%, 5 +/- 4%, 2 +/- 2%, and 1 +/- 2% from one to four parameters, respectively. More than three parameters were associated with a specificity above 90% and a positive predictive value above 65%. Reproducibility of this global strategy was 91%.

Conclusion: A multiparametric echocardiographic strategy based on the association of conventional criteria is a better indicator of CRT response than the existing single parametric approaches.

Publication types

  • Multicenter Study
  • Validation Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Algorithms*
  • Cardiac Pacing, Artificial*
  • Epidemiologic Methods
  • Female
  • Heart Failure / complications
  • Heart Failure / diagnostic imaging*
  • Heart Failure / therapy*
  • Humans
  • Male
  • Middle Aged
  • Stroke Volume
  • Ultrasonography
  • Ventricular Dysfunction, Left / complications
  • Ventricular Dysfunction, Left / diagnostic imaging*
  • Ventricular Dysfunction, Left / therapy*