Objectives: The Impella recovery axial-flow system is a mechanical assist system for use in acute heart failure. This retrospective study reports the use of the device at three cardiothoracic units in Sweden.
Design: Fifty patients (35 men, mean age 55.8 years, range 26 to 84 years) underwent implantation of 26 ImpellaLP 2.5/5.0 (support-time 0.1 to 14 days), 16 ImpellaLD (support-time 1 to 7 days) and 8 ImpellaRD (support-time 0.1 to 8 days) between 2003 and 2007. Implantation was performed because of postcardiotomy heart failure (surgical group, n=33) or for various states of heart failure in cardiological patients (non-surgical group, n=17). The intention for the treatments was mainly to use the pump as a "bridge-to-recovery".
Results: Early mortality in the surgical and non-surgical groups was 45% and 23%, respectively. Complications included infection, 36% and right ventricular failure, 28%. Cardiac output and cardiac power output postoperatively were significantly higher among survivors than non-survivors.
Conclusions: The Impella recovery axial-flow system facilitates treatment in acute heart failure. Early intervention in patients with acute heart failure and optimized hemodynamics in the post-implantation period seem to be of importance for long-term survival. Insufficient early response to therapy should urge to consider further treatment options.