A low-level gamma-ray spectrometry system, based on an HPGe-detector with 92% relative efficiency recently installed in the underground laboratory Felsenkeller at 110 m water equivalent (w.e.) depth, is described. The integral background count rate normalised to the Ge-crystal mass in the energy range from 40 keV until 2.7 MeV of 0.034 s(-1)kg(-1) has been achieved by careful material selection of the detector construction material, a graded shielding construction and effective radon suppression. The detector is highly suitable for the effective surveillance of water for human consumption with decision thresholds for (226)Ra and (228)Ra in the order of some mBq L(-1).