Validation of a bedside activated clotting time test (Hemochron Jr II Signature) with low dose heparin therapy

Anaesthesia. 2009 Apr;64(4):430-4. doi: 10.1111/j.1365-2044.2008.05822.x.

Abstract

The purpose of this study was to validate a device designed to measure activated clotting time in low-range heparin plasma concentrations (ACT-LR) prospectively during the post-operative period of vascular surgery. Measurement of ACT-LR and activated partial thromboplastin time (APTT) were performed before heparinisation (T0) and at the end of surgery (T1). ACT-LR(T1) and DeltaACT-LR (defined as ACT-LR(T1) - ACT-LR(T0)) were evaluated as diagnostic tests for excessive anticoagulation, defined by APTT more than twice the laboratory's normal, by Bland-Altman method and receiver operating characteristic (ROC) curves. In 103 patients, mean (SD) ACT-LR was 137 (33) s at T0 and 176 (39) s at T1. Bland-Altman graph did not show a good agreement between APTT and ACT-LR. Areas under ROC curves were 0.82 (95% CI: 0.75-0.89) and 0.87 (95% CI: 0.80-0.93) for ACT-LR(T1) and DeltaACT-LR, respectively. Using a threshold of 32 s for DeltaACT-LR, test sensitivity was 87% (95% CI: 81-93%), specificity was 85% (95% CI: 78-92%), positive predictive value was 90% (95% CI: 84-96%) and negative predictive value was 81% (95% CI: 73-86%). While DeltaACT-LR may have some potential in evaluating excessive anticoagulation in vascular surgery, the poor correlation between ACT-LR and APTT does not support its routine use.

Publication types

  • Validation Study

MeSH terms

  • Aged
  • Anticoagulants / administration & dosage*
  • Blood Coagulation Tests / methods*
  • Drug Administration Schedule
  • Drug Monitoring / methods*
  • Epidemiologic Methods
  • Female
  • Heparin / administration & dosage*
  • Humans
  • Male
  • Middle Aged
  • Partial Thromboplastin Time
  • Point-of-Care Systems*
  • Postoperative Care / methods
  • Vascular Surgical Procedures

Substances

  • Anticoagulants
  • Heparin