Performance of single-photon-counting PILATUS detector modules

J Synchrotron Radiat. 2009 May;16(Pt 3):368-75. doi: 10.1107/S0909049509009911. Epub 2009 Apr 7.

Abstract

PILATUS is a silicon hybrid pixel detector system, operating in single-photon-counting mode, that has been developed at the Paul Scherrer Institut for the needs of macromolecular crystallography at the Swiss Light Source (SLS). A calibrated PILATUS module has been characterized with monochromatic synchrotron radiation. The influence of charge sharing on the count rate and the overall energy resolution of the detector were investigated. The dead-time of the system was determined using the attenuated direct synchrotron beam. A single module detector was also tested in surface diffraction experiments at the SLS, whereby its performance regarding fluorescence suppression and saturation tolerance were evaluated, and have shown to greatly improve the sensitivity, reliability and speed of surface diffraction data acquisition.

MeSH terms

  • Computer-Aided Design
  • Equipment Design
  • Equipment Failure Analysis
  • Photometry / instrumentation*
  • Photometry / methods
  • Photons
  • Radiation Dosage
  • Radiometry / instrumentation*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Transducers*
  • X-Ray Diffraction / instrumentation*