We investigated the diffuse optical image errors resulting from systematic errors in the background scattering and absorption coefficients, Gaussian noise in the measurements, and the depth at which the image is reconstructed when using a 2D linear reconstruction algorithm for a 3D object. The fourth Born perturbation approach was used to generate reflectance measurements and k-space tomography was used for the reconstruction. Our simulations using both single and dual wavelengths show large systematic errors in the absolute reconstructed absorption coefficients and corresponding hemoglobin concentrations, while the errors in the relative oxy- and deoxy- hemoglobin concentrations are acceptable. The greatest difference arises from a systematic error in the depth at which an image is reconstructed. While an absolute reconstruction of the hemoglobin concentrations can deviate by 100% for a depth error of ñ1 mm, the error in the relative concentrations is less than 5%. These results demonstrate that while quantitative diffuse optical tomography is difficult, images of the relative concentrations of oxy- and deoxy-hemoglobin are accurate and robust. Other results, not presented, confirm that these findings hold for other linear reconstruction techniques (i.e. SVD and SIRT) as well as for transmission through slab geometries.