A rapid and reliable method for the quantitation of tricyclic antidepressants in serum using HPLC-MS/MS

Clin Biochem. 2009 Aug;42(12):1300-7. doi: 10.1016/j.clinbiochem.2009.05.006. Epub 2009 May 20.

Abstract

Objective: Here we describe a liquid chromatography-tandem mass spectrometry method for the detection of tricyclic antidepressant drugs (amitriptyline, desipramine, imipramine, and nortriptyline) in serum.

Design and methods: Following a protein precipitation from serum and dilution of supernatant with water, these analytes and their internal standards (deuterated versions of each drug) were injected, separated, and eluted from a Hypersil Gold C-18 (50x2.1 mm) analytical column with a gradient of water and acetonitrile each with 0.1% formic acid. Analytes were then ionized and detected over a 3.5 minute analysis time by electrospray ionization mass spectrometry with multiple reaction monitoring. Matrix effects were characterized using three methods: post-column infusion, pre- and post-extraction addition of analyte to matrix, and comparison of neet to serum preparations.

Results: The inter-day and intra-day coefficients of variation for all compounds were < or =12%. The limit of detection for all drugs was <15 ng/mL and the limit quantitation for all drugs was <22 ng/mL. Recoveries were between 97 and 131% for all drugs. Patient method comparison and proficiency samples were run with acceptable results.

Conclusion: We have developed a specific and sensitive method for rapid quantitation of tricyclic antidepressants in serum, suitable for use in the clinical laboratory.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Antidepressive Agents, Tricyclic / blood*
  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tandem Mass Spectrometry / methods*
  • Time Factors

Substances

  • Antidepressive Agents, Tricyclic