Accuracy and interobserver reliability of three-dimensional rotational angiography versus mathematical models for volumetric measurement of intracranial aneurysms

J Clin Neurosci. 2009 Sep;16(9):1195-8. doi: 10.1016/j.jocn.2008.11.011. Epub 2009 Jun 3.

Abstract

We aimed to compare the inter-rater reliability relating to the volumetric measurement of intracranial aneurysms obtained with three-dimensional rotational angiography (3D-RA) compared with two commonly used mathematical models. Ten randomly selected aneurysms were measured by 3 independent assessors using 3D-RA and 2 mathematical models for the calculation of an ellipsoid: Equation (1), Vab=4/3pi x (a/2)x(b/2)x(a + b/4); and Equation (2), Vabc=(4/3)pi x (a/2)x(b/2)x(c/2). The inter-rater reliability for each method was: 3D-RA, 0.99; Vabc, 0.90; and Vab, 0.89. The 95% confidence interval for the mean difference between 3D-RA and Vabc was not significantly different, whereas there was a significant difference between 3D-RA and Vab. Vab gave consistently higher estimates than 3D-RA. This was especially true for aneurysms with larger volumes. The use of 3D-RA to undertake volumetric measurements of intra-cranial aneurysms is both valid and reproducible for different assessors.

MeSH terms

  • Algorithms
  • Cerebral Angiography*
  • Humans
  • Intracranial Aneurysm / pathology*
  • Magnetic Resonance Angiography
  • Models, Statistical
  • Observer Variation
  • Reproducibility of Results
  • Retrospective Studies