The authors modified Gianturco expandable metallic stents to make them relocatable. Two tandem designs were made. The first design had long struts and, thus, more evenly distributed expansive force; the second had short struts and more flexibility and was better suited for use in curved strictures. Both designs (two to four in tandem, 0.012-, 0.014-, and 0.018-inch wire) were tested successfully in vitro, and both designs (two to four in tandem, 0.012-inch wire) were successfully placed, retrieved, and relocated in the inferior vena cava of five dogs.