Testing efficiency transfer codes for equivalence

Appl Radiat Isot. 2010 Feb;68(2):355-9. doi: 10.1016/j.apradiso.2009.10.012. Epub 2009 Oct 17.

Abstract

Four general Monte Carlo codes (GEANT3, PENELOPE, MCNP and EGS4) and five dedicated packages for efficiency determination in gamma-ray spectrometry (ANGLE, DETEFF, GESPECOR, ETNA and EFFTRAN) were checked for equivalence by applying them to the calculation of efficiency transfer (ET) factors for a set of well-defined sample parameters, detector parameters and energies typically encountered in environmental radioactivity measurements. The differences between the results of the different codes never exceeded a few percent and were lower than 2% in the majority of cases.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Equipment Design
  • Equipment Failure Analysis
  • Radiometry / instrumentation*
  • Radiometry / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Software*