The geometric accuracy of frameless stereotactic radiosurgery using a 6D robotic couch system

Phys Med Biol. 2010 Jan 7;55(1):1-10. doi: 10.1088/0031-9155/55/1/001.

Abstract

The aim of this paper is to assess the overall geometric accuracy of the Novalis system using the Robotic Tilt Module in terms of the uncertainty in frameless stereotactic radiotherapy. We analyzed the following three metrics: (1) the correction accuracy of the robotic couch, (2) the uncertainty of the isocenter position with gantry and couch rotation, and (3) the shift in position between the isocenter and central point detected with the ExacTrac x-ray system. Based on the concept of uncertainty, the overall accuracy was calculated from these values. The accuracy in positional correction with the robotic couch was 0.07 +/- 0.22 mm, the positional shift of the isocenter associated with gantry rotation was 0.35 mm, the positional shift of the isocenter associated with couch rotation was 0.38 mm and the difference in position between the isocenter and the ExacTrac x-ray system was 0.30 mm. The accuracy of intracranial stereotactic radiosurgery with the Novalis system in our clinic was 0.31 +/- 0.77 mm. The overall geometric accuracy based on the concept of uncertainty was 0.31 +/- 0.77 mm, which is within the tolerance given in the American Association of Physicists in Medicine report no. 54.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Brain / surgery
  • Humans
  • Models, Biological
  • Patient Positioning
  • Phantoms, Imaging
  • Radiosurgery / instrumentation*
  • Reproducibility of Results
  • Robotics / instrumentation*
  • Uncertainty