Development of a new lead for in utero fetal pacing

Prenat Diagn. 2010 Feb;30(2):122-6. doi: 10.1002/pd.2422.

Abstract

Background: The results from preliminary studies on fetal pacing preclude an application to the human fetus. The purpose of this study was to evaluate the feasibility of acute fetal cardiac pacing following ultrasound-guided epicardial anchorage of a new pacing lead dedicated to the fetal heart.

Methods: In 12 pregnant ewes, a needle was placed in contact with cardiac chambers under echo-guidance. Using the lumen of the needle, a 3-Fr pacing lead was advanced and fixed on the cardiac wall using a built-in screw. Pacing was attempted using an external pacing system.

Results: Lead insertion was successful in all but one. In the first fetus, the needle was advanced too far into the right ventricle leading to pericardial bleeding. The fetal heart rate was overdriven in all at 220/ min. After assessment, pacing lead was unscrewed and retrieved without any complication. Pregnancy was electively interrupted in six animals. The other six fetuses were delivered vaginally after uneventful pregnancy and autopsy at day 3 did not show any significant injury.

Conclusions: Transcutaneous echo-guided fetal acute pacing is feasible using a specially designed lead. Additional refinement of both the equipments and the operative technique are needed before considering permanent pacing in the human fetus.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Cardiac Pacing, Artificial / methods*
  • Electrodes
  • Electrodes, Implanted
  • Female
  • Fetal Heart / physiology*
  • Pacemaker, Artificial*
  • Pregnancy
  • Sheep