Background: The VIO soft-coagulation system (SCS) is a new device for tissue coagulation. The current study evaluated the efficacy of the SCS when used for liver resection.
Methods: The 252 patients were divided into 2 groups; in 155 patients (conventional group), liver transection was performed using an ultrasonic dissector and saline-coupled bipolar electrocautery for hemostasis. In 97 patients (SCS group), the SCS was used instead of bipolar electrocautery.
Results: The median blood loss and surgical time were less in the SCS group than in the conventional group (350 vs 640 mL, P = .0028; 280 vs 398 min, P < .0001). No significant differences were found in postoperative complications between the SCS group (32.0%) and the conventional group (40.6%). The risk factors for bleeding were nonuse of the SCS (P = .0039), macroscopic vascular invasion of the hepatic tumors (P = .0088), and collagen type IV value in the sera >200 (P = .0250) on multivariate analysis. In a subgroup analysis, in the collagen type IV value >200 subgroup, the tumor diameter >5 cm subgroup, and the inflow nonocclusion subgroup, use of the SCS decreased surgical bleeding (P = .0120, P = .0126, and P = .0032, respectively) and surgical time (P = .0001, P < .0001, and P = .0036, respectively) compared with the conventional group. Furthermore, even in the major hepatectomy group, the SCS use decreased surgical time (P < .0001).
Conclusion: The SCS is an effective and safe device for decreasing surgical time and surgical bleeding without increasing the rate of bile leakage and causing other complications.
Copyright © 2011 Mosby, Inc. All rights reserved.