Purpose/objective(s): Volumetric modulated arc therapy (VMAT) allows for intensity-modulated radiation delivery during gantry rotation with dynamic MLC motion, variable dose rates and gantry speed modulation. We compared VMAT plans with 3D-CRT for hypofractionated lung radiotherapy.
Materials/methods: Twenty-one 3D-CRT plans for Stage IA lung cancer previously treated stereotactically were selected. VMAT plans were generated by optimizing machine aperture shape and radiation intensity at 10 degrees intervals. A partial arc range of 180 degrees was manually selected to coincide with tumor location. The arc was resampled down to 5 degrees intervals to ensure dose calculation accuracy. Identical planning objectives were used for VMAT/3D-CRT. Parameters assessed included dose to PTV and organs-at-risk (OAR), monitor units, and multiple conformity and homogeneity indices. Plans were delivered to a phantom for time comparison.
Results: Lung V(20/12.5/10/5) were less with VMAT (relative reduction 4.5%, p = .02; 3.2%, p = .01; 2.6%, p = .01; 4.2%, p = .03, respectively). Mean/maximum-doses to PTV, dose to additional OARs, 95% isodose line conformity, and target volume homogeneity were equivalent. VMAT improved conformity at both the 80% (1.87 vs. 1.93, p = .08) and 50% isodose lines (5.19 vs. 5.65, p = .01). Treatment times were reduced significantly with VMAT (mean 6.1 vs. 11.9 min, p < .01).
Conclusions: Single arc VMAT planning achieves highly conformal dose distributions while controlling dose to critical structures, including significant reduction in lung dose volume parameters. Employing a VMAT technique decreases treatment times by 37-63%, reducing the chance of error introduced by intrafraction variation. The quality and efficiency of VMAT is ideally suited for stereotactic lung radiotherapy delivery.
Copyright 2010 Elsevier Ireland Ltd. All rights reserved.