Technical validation of an autoantibody test for lung cancer

Ann Oncol. 2010 Aug;21(8):1687-1693. doi: 10.1093/annonc/mdp606. Epub 2010 Feb 2.

Abstract

Background: Publications on autoantibodies to tumour-associated antigens (TAAs) have failed to show either calibration or reproducibility data. The validation of a panel of six TAAs to which autoantibodies have been described is reported here.

Materials and methods: Three separate groups of patients with newly diagnosed lung cancer were identified, along with control individuals, and their samples used to validate an enzyme-linked immunosorbant assay. Precision, linearity, assay reproducibility and antigen batch reproducibility were all assessed.

Results: For between-replicate error, samples with higher signals gave coefficients of variation (CVs) in the range 7%-15%. CVs for between-plate variation were only 1%-2% higher. For between-run error, CVs were in the range 15%-28%. In linearity studies, the slope was close to 1.0 and correlation coefficient values were generally >0.8. The sensitivity and specificity of individual batches of antigen varied slightly between groups of patients; however, the sensitivity and specificity of the panel of antigens as a whole remained constant. The validity of the calibration system was demonstrated.

Conclusions: A calibrated six-panel assay of TAAs has been validated for identifying nearly 40% of primary lung cancers via a peripheral blood test. Levels of reproducibility, precision and linearity would be acceptable for an assay used in a regulated clinical setting.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Autoantibodies / blood*
  • Case-Control Studies
  • Enzyme-Linked Immunosorbent Assay / methods*
  • Humans
  • Lung Neoplasms / immunology*
  • Middle Aged
  • Quality Control
  • Reproducibility of Results