Homogentisic acid interference in routine urine creatinine determination

Mol Genet Metab. 2010 May;100(1):103-4. doi: 10.1016/j.ymgme.2010.01.006. Epub 2010 Jan 21.

Abstract

We report the artifactual elevation of homogentisic acid (HGA) in urine from alkaptonuric patients after replacing the creatinine method (Jaffe reaction) in our laboratory with an automated enzymatic method. Samples with elevated HGA by GC-MS had lower creatinine values as determined by the enzymatic method than by the Jaffe reaction. The low creatinine values were due to interference by HGA in the enzymatic method. The enzymatic method is unsuitable for creatinine determination in urine of patients with alkaptonuria.