Accuracy of the SEVEN continuous glucose monitoring system: comparison with frequently sampled venous glucose measurements

J Diabetes Sci Technol. 2009 Sep 1;3(5):1146-54. doi: 10.1177/193229680900300519.

Abstract

Background: The purpose of this study was to compare the accuracy of measurements obtained from the DexCom SEVEN system with Yellow Springs Instrument (YSI) laboratory measurements of venous blood glucose.

Methods: Seventy-two subjects with insulin-requiring diabetes, aged 18-71, were enrolled in a multicenter, prospective single-arm study. All participants wore the SEVEN continuous glucose monitoring (CGM) system for one, 7-day wear period. Calibration with capillary finger stick measurements was performed 2 hours after sensor insertion and once every 12 hours thereafter. A subset of subjects (28) wore two systems simultaneously to assess precision. All subjects participated in one, 10-hour in-clinic session on day 1, 4, or 7 of the study to compare CGM measurements against a laboratory method (YSI analyzer) using venous measurements taken once every 20 minutes. Carbohydrate consumption and insulin dosing were adjusted in order to obtain a broad range of glucose values.

Results: Comparison of CGM measurements with the laboratory reference method (n = 2318) gave mean and median absolute relative differences (ARDs) of 16.7 and 13.2%, respectively. The percentage was 70.4% in the clinically accurate Clarke error grid A zone and 27.5% in the benign error B zone. Performance of the SEVEN system was consistent over time with mean and median ARD lowest on day 7 as compared to YSI (13.3 and 10.2%, respectively). Average sensor time lag was 5 minutes.

Conclusions: Measurements of the DexCom SEVEN system were found to be consistent and accurate compared with venous measurements made using a laboratory reference method over 7 days of wear.

Publication types

  • Clinical Trial
  • Comment
  • Comparative Study
  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Blood Glucose / metabolism*
  • Blood Glucose Self-Monitoring / instrumentation*
  • Blood Glucose Self-Monitoring / standards
  • Blood Specimen Collection
  • Calibration
  • Catheterization, Peripheral
  • Diabetes Mellitus, Type 1 / blood
  • Diabetes Mellitus, Type 1 / diagnosis*
  • Diabetes Mellitus, Type 1 / drug therapy
  • Diabetes Mellitus, Type 2 / blood
  • Diabetes Mellitus, Type 2 / diagnosis*
  • Diabetes Mellitus, Type 2 / drug therapy
  • Diagnostic Equipment*
  • Dietary Carbohydrates / administration & dosage
  • Dietary Carbohydrates / metabolism
  • Equipment Design
  • Female
  • Humans
  • Hypoglycemic Agents / administration & dosage
  • Insulin / administration & dosage
  • Male
  • Middle Aged
  • Predictive Value of Tests
  • Prospective Studies
  • Reproducibility of Results
  • Time Factors
  • United States
  • Young Adult

Substances

  • Blood Glucose
  • Dietary Carbohydrates
  • Hypoglycemic Agents
  • Insulin