Quality control of micro-computed tomography systems

Radiat Prot Dosimetry. 2010 Apr-May;139(1-3):463-7. doi: 10.1093/rpd/ncq088. Epub 2010 Mar 11.

Abstract

The rapid proliferation of micro-computed tomography (micro-CT) scanners in preclinical small animal studies has created a need for a method on scanner performance evaluation and scan parameter optimisation. The purpose of this study was to investigate the performance of the scanner with a dedicated micro-CT phantom. The phantom was developed with different independent sections that allow for measurement of major scanner characteristics such as uniformity, linearity, contrast response, dosimetry and resolution. The results of a thorough investigation are discussed.

MeSH terms

  • Equipment Failure Analysis
  • Phantoms, Imaging
  • Quality Assurance, Health Care / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • X-Ray Microtomography / instrumentation*
  • X-Ray Microtomography / methods*