The sensitivity of direct identification from positive BacT/ALERT™ (bioMérieux) blood culture bottles by matrix-assisted laser desorption ionization time-of-flight mass spectrometry is low

Clin Microbiol Infect. 2011 Feb;17(2):192-5. doi: 10.1111/j.1469-0691.2010.03229.x.

Abstract

Recently, matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) has been presented as a novel method for the direct identification of bacteria from positive blood culture bottles. The rate of the MALDI TOF MS-based identification in the present study from positive BacT/ALERT (bioMérieux, Marcy l'Etoile, France) blood culture bottles was 30%, which is far below the previously reported sensitivities using the BACTEC (Becton Dickinson, Franklin Lakes, NJ, USA) system. We also found evidence that the Biotyper algorithm did not identify a second pathogen in cases of positive BacT/ALERT blood culture bottles containing two different species.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Bacteremia / diagnosis*
  • Bacteria / isolation & purification*
  • Bacteriological Techniques / methods*
  • Blood / microbiology*
  • Humans
  • Sensitivity and Specificity
  • Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization / methods*