Digital instrumentation and management of dead time: first results on a NaI well-type detector setup

Appl Radiat Isot. 2010 Jul-Aug;68(7-8):1314-8; discussion 1318-9. doi: 10.1016/j.apradiso.2010.01.040. Epub 2010 Feb 12.

Abstract

The LNE-LNHB is engaged in a development program on digital instrumentation, the first step being the instrumentation of a NaI well-type detector set-up. The prototype acquisition card and its technical specifications are presented together with the first comparison with the classical NIM-based acquisition chain, for counting rates up to 100 kcps. The digital instrumentation is shown to be counting-loss free in this range. This validates the main option adopted in this project, namely the implementation of an extending dead time with live-time measurement already successfully used in the MTR2 NIM module developed at LNE-LNHB.