X-ray fluorescence (XRF) set-up with a low power X-ray tube

Appl Radiat Isot. 2010 Oct;68(10):1922-7. doi: 10.1016/j.apradiso.2010.05.001. Epub 2010 May 12.

Abstract

The X-ray fluorescence set-up with a 100 W X-ray tube comprises a computer controlled system developed for remote operation and monitoring of tube and an adjustable stable 3D arrangement to procure variable excitation energies with low scattered background. The system was tested at different filament currents/anode voltages. The MDL of the set-up at 0.05-1.00 mA/4-12 kV is found approximately (1-100)ppm for K and L excitations and approximately (200-700)ppm for M excitations of elements and improves with filament current and anode voltage. Moreover, L measurements for Sm and Eu at five K X-ray energies of elements(Z=29-40) and analytical determination in some synthetic samples were undertaken.