An investigation into the bias between liquid chromatography-tandem mass spectrometry and an enzyme multiplied immunoassay technique for the measurement of mycophenolic acid

Ther Drug Monit. 2010 Aug;32(4):420-6. doi: 10.1097/FTD.0b013e3181e6b348.

Abstract

Mycophenolic acid is now the second most widely used immunosuppressant in solid organ transplantation. Overestimation of mycophenolic acid concentration is a recognized problem of immunoassay, and high-performance liquid chromatography with ultraviolet detection methods have long analysis times and a risk of analyte coelution which may compromise high sample throughput in a clinically meaningful time frame. A novel liquid chromatography-tandem mass spectrometry assay for mycophenolic acid was developed using very small (10 microL) sample volumes and evaluated in comparison with an established immunological assay. The enzyme mediated immunoassay showed a median positive bias compared with liquid chromatography-tandem mass spectrometry of 14.6%. Linear regression analysis showed a significant positive impact of bilirubin (r2 = 0.230) on bias with further increases of r2 to 0.261, 0.286, and 0.294 with the stepwise addition of creatinine, hematocrit, and gamma-glutamyl transpeptidase, respectively. The impact of comedication and transplant type depended on the patient population: analysis of all samples showed opposing effects to analysis of those samples lacking data with biochemical variables above. The liquid chromatography-tandem mass spectrometry method described in this report is capable of measuring mycophenolic acid concentrations in very small sample volumes and in a timely fashion without the significant overestimates characterizing enzyme mediated immunoassay measurements in patients with serologic features characterizing liver or renal graft rejection.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Bias
  • Chromatography, High Pressure Liquid / statistics & numerical data*
  • Drug Interactions
  • Drug Monitoring / methods
  • Drug Monitoring / standards
  • Enzyme Multiplied Immunoassay Technique / statistics & numerical data*
  • Female
  • Humans
  • Immunosuppressive Agents / blood*
  • Kidney Function Tests
  • Linear Models
  • Male
  • Mycophenolic Acid / blood*
  • Population
  • Quality Control
  • Tandem Mass Spectrometry / statistics & numerical data*

Substances

  • Immunosuppressive Agents
  • Mycophenolic Acid