Thermoluminescence dosimetry for skin dose assessment during intraoperative radiotherapy for early breast cancer

Australas Phys Eng Sci Med. 2010 Jun;33(2):211-4. doi: 10.1007/s13246-010-0019-3. Epub 2010 Jul 9.

Abstract

Dosimetry for intraoperative radiotherapy (IORT) after wide local excision for breast cancer using a 50 kV X-ray needle (Intrabeam) was performed in vivo using thermoluminescence dosimetry. Eight LiF:Mg,Ti chips were placed on the skin around the incision site after wide local excision while the tumour bed was irradiated to a prescribed dose of 5 Gy 10 mm from the applicator surface. The maximum and mean measured skin dose for 57 patients ranged from 0.64 to 7.1 Gy and 0.56 to 4.78 Gy, respectively, reflecting different tissue thicknesses overlying the applicator. The average maximum dose of 2.93+/-1.46 Gy was below the threshold for severe radiation skin toxicity.